Maintenance and Inspections
4.
Plug in the AC power cable.
• The On Battery Alarm stops.
• The AC power lamp is lit.
• The AC power indicator changes to
.
5.
Change the motor speed to 0 RPM to stop the Drive Motor.
6. Turn
off
the
power.
Inspection method of the Drive Motor Disconnected Alarm function and the ABD/Flow
Sensor Disconnected Alarm function (Before each use)
1.
Disconnect the plug of the Drive Motor and turn on the
power.
• “Drive Motor Disconnected” is displayed on the message area and an alarm
sounds. (Figure 5-2)
2. Tap .
(Figure
5-3)
• The alarm is muted.
Figure 5-2
3.
Turn off the power and connect the Drive Motor.
4.
Disconnect the plug of the ABD/Flow Sensor and turn on
the power.
Figure 5-3
• “ABD/Flow Sensor Disconnected” is displayed on the message area and an alarm
sounds. (Figure 5-4)
5. Tap .
(Figure
5-5)
• The alarm is muted.
Figure 5-4
6.
Turn off the power and connect the ABD/Flow Sensor.
Figure 5-5
124
Check
Checks
Check
Check
Check