Instructions for use
Thoroughly read “Warnings/Precautions” (Page 4 - 5) before use.
Set Up
1. Set a holder and a drive motor.
・
Fix the drive motor to the holder tightly with screws for
the drive motor (Figure 1).
2. Fix the holder to the pole with a pole clamp of the holder
(Figure 2).
3. Take the product out of the unit package and confirm there is
nothing wrong.
・
Do not use if the unit package or the product has
been damaged or soiled.
・
When using CX
*
XSA0171, CX
*
XSA0173
or CX
*
XSA0251, do not hold the product by
the tube between oxygenator and centrifugal
pump.
This may cause heavy load at the connection
parts of tubing and lead to leakage.
4. Confirm that all the screws of 3-way stopcocks, caps and
connectors are fixed tightly, and that the clamps of the arterial
line (red), the venous line (blue) and the priming line are
open.
5. Set the centrifugal pump to the drive motor fixed to the holder.
(For CX
*
XSA: Remove the cover of the centrifugal pump.)
①
Remove the magnet cover from the drive motor (Figure 3).
②
Insert the rim of the centrifugal pump into the fixed hook of
the drive motor (Figure 4).
③
Insert the rim of back side of the oxygenator into the fixed
hook of the holder (XX
*
EB04), and push forward and set the
oxygenator (Figure 5).
Instructions for Use
CAUTION
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...