Other Operation Procedures
Setup of the Controller
1.
Turn the air bubble detector function on.
For details about how to set, refer to “Setting the Air Bubble Detector
Function” (page 104).
2.
Set the safety connection of the Air Bubble Detected Alarm
to “Pump Coast”.
For details about how to set, refer to “Setting the Safety Connection”
(page 91).
• When activating the safety connection, be sure to set the air bubble detector function on. The safety connection is not
activated without the air bubble detector function.
Actions after activating the safety connection
1.
When the Air Bubble Detected Alarm is issued, the motor
speed is lowered to the coast speed.
The status lamp starts to flash in red and the alarm sounds, while “Air Bubble
Detected Alarm” and “Air Bubble: Pump Coasting” are displayed on the
message area. (Figure 2-2)
• When the motor speed is equal to or less than the coast speed, the safety
connection is not activated.
2.
Remove the air bubble from the blood circuit.
For details about how to handle, refer to “Alarms and Troubleshooting”
(page 132).
Figure 2-2
3. Tap
[Air
Bubble
Detected
Alarm] on the message area,
and after confirming the displayed message, tap [OK].
(Figure 2-3)
The Air Bubble Detected Alarm is cleared, and “Air Bubble Detected Alarm”
on the message area is turned off.
• The Air Bubble Detected Alarm can also be cleared by performing the step 4.
Figure 2-3
67
b
When the safety connection of the Air Bubble Detected Alarm is set to “Pump Coast”
Note
Note
Note
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...