Specifications
Alarms and safety functions
Air Bubble Detected Alarm, Flow Alarm, Back Flow Alarm, Pressure
Alarm (High priority), Pressure Alarm (Low priority), Temperature
Alarm, Low Battery Alarm, Battery Empty Alarm, Battery-powered
operation started Alarm, On Battery Alarm, Serious System Error Alarm
(High priority), System Error Alarm (Medium priority), CDI Connection
Error / LAN Connection Error Alarm, Drive Motor Disconnected
Alarm, ABD/Flow Sensor Disconnected Alarm, Flow Signal Unstable
Alarm, Pressure Sensor Disconnected Alarm, Temperature Sensor
Disconnected Alarm, Clamp Closed Alarm, Clamp Occlusion
Error Alarm, Clamp Error Alarm, Clamp Not Ready Alarm, Clamp
Disconnected Alarm, Re-alarm
Available functions
Perfusion index calculation function, Count up timer function, Count
down timer function, Key lock function, History function, External
communication function, Date and time setting, Language setting,
Volume setting, LCD touch panel display brightness setting, Pressure
unit setup, Admin setup, Alarm/safety apparatus operation setting reset
function
Use conditions
Ambient temperature: 10 to 40°C
Relative humidity:
30 to 85% RH (non-condensing)
Pressure:
80 to 106 kPa
Storage conditions or transport conditions
Ambient temperature: -20 to 45°C
Relative humidity:
10 to 95% RH (non-condensing)
Pressure: 70
to
106
kPa
Power supply
AC100-240V 50-60
Hz
Battery (lithium ion battery)
• Run time:
One hour or more (running
(
*
)
, when fully charged)
• Charging time: Within six hours (While power is OFF or running
(
*
)
with AC power connected)
*: Motor speed 2500 RPM, flow 4.00 LPM (L/min), apparatus operated
with all sensors on, ambient temperature 25°C, new battery
Power consumption
250 VA (Steady-state)
400 VA (Peak: When the motor is rapidly accelerating, auto-priming is
operating)
Fuse
Time-lag fuse (Type and rated voltage: T8AH250 V)
Voltage rating: 250 V
Current rating: 8 A
Interrupt rating: 1500 A at 250 V
Classification
Electric shock protection
Class I device and internal power source device
Defibrillation-proof type CF applied part (Pressure sensor, temperature
sensor)
The classification for use with
flammable anaesthetics
Device not suitable for use around flammable gas
Mode of operation
Continuous operation
Protection against harmful
ingress of water or particulate
matter
IP42 (Controller)
IP54 (Fast Clamp)
Dimensions,
unit weight
Controller
Dimensions
160 mm (width) x 300 mm (height) x 330 mm (depth)
* Excluding handle, rubber feet, and protrusions
Unit weight
8 kg
Drive Motor
Cable length
3 m
Unit weight
2.6 kg
ABD/Flow Sensor
Cable length
2.8 m
Fast Clamp
Cable length
1.8 m
Unit weight
1.7 kg
139
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...