Other Operation Procedures
2.
Clamp the blood line of the patient side of the Fast Clamp
with forceps. (Figure 2-5)
Figure 2-5
3.
Remove the air bubble from the blood circuit.
For details about how to handle, refer to “Alarms and Troubleshooting”
(page 132).
4. Tap
[Air
Bubble
Detected
Alarm] on the message area,
and after confirming the displayed message, tap [OK].
(Figure 2-6)
The Air Bubble Detected Alarm is cleared, and “Air Bubble Detected Alarm”
on the message area is turned off.
• The Air Bubble Detected Alarm can also be cleared by performing the step 5 or
step 6.
Figure 2-6
5.
Pull up the lever of the Fast Clamp until you hear a click,
and remove the occlusion with the Fast Clamp. (Figure 2-7)
“Air Bubble: Clamp Close” on the message area is turned off.
• After pulling up the lever, the lever has returned to the original position.
• If the lever does not return to the original position, pull down the lever. If it does
not return to the original position, a failure may have occurred.
Stop using the system immediately, and contact Terumo’s service representative.
Figure 2-7
69
Note
Check
Note
Blood line
Fast Clamp
Forceps
Lever
S
ide
of
the pa
tie
nt
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...