Various menu settings
5.
Tap [OK] after confirming the message displayed.
(Figure 3-65)
The password is changed.
Figure 3-65
Tap the icon of the menu you want to set, and then the Admin Setup screen opens.
Check the corresponding item for the setup method of each menu.
Icon name
Function
Factory setting
Remark
Reference
page
Flow
Preset
Sets the admin setting of the
Flow Alarm threshold.
Alarm upper limit: 7.0 LPM (L/min)
Alarm lower limit: 0.5 LPM (L/min)
The setting will take
effect after restarting.*
page 102
Pressure
Preset
Sets the on/off status of the
pressure measuring function
and the admin setting of the
threshold for the Pressure
alarm and Pressure Alert for
every channel.
The display unit can also be
set.
Alarm upper limit: 360 mmHg
Alert upper limit: 300 mmHg
Alert lower limit: 70 mmHg
Alarm lower limit: 50 mmHg
The setting will take
effect after restarting.*
page 103
On/Off: On
mmHg/kPa: mmHg
The setting will take
effect instantly.
Temperature
Preset
Sets the on/off status of
the temperature measuring
function and the admin
setting of the threshold value
for the Temperature Alert for
every channel.
Upper limit temperature: 42.0°C
Lower limit temperature: 30.0°C
The setting will take
effect after restarting.*
page 104
On/Off: On
Fast
clamp
Sets the on/off status for the
admin setting of the Fast
Clamp.
On/Off: Off
The setting will take
effect after restarting.*
page 104
Air Bubble
Sets the on/off status of the
air bubble detector function
and the admin setting of the
detection sensitivity.
On/Off: On
The setting will take
effect after restarting.*
page 104
Detection sensitivity (High/
Medium/Low): Low
The setting will take
effect instantly.
101
Admin Setup Screen
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...