Operation Procedure
1.
Take out the centrifugal pump from the packaging and
confirm that there are no abnormalities.
2.
Connect the extracorporeal circulation circuit line to the
blood inlet port and the blood outlet port of the centrifugal
pump. (Figure 1-30)
Figure 1-30
1.
Remove the magnet cover of the Drive Motor. (Figure 1-31)
Pull the slide hook of the Drive Motor, and remove the magnet cover at an
angle.
Figure 1-31
2.
Attach the centrifugal pump to the pump receptacle of the
Drive Motor. (Figure 1-32)
Insert the rim of the centrifugal pump into the fixed hook of the Drive Motor
and press the pump against the slide hook to attach.
• The centrifugal pump is securely attached with the slide hook and the fixed hook.
Figure 1-32
43
Centrifugal Pump Connection to the Tube Pack
• Do not use this system with any centrifugal pump other than the dedicated centrifugal pump (CAPIOX centrifugal pump,
CAPIOX centrifugal pump as the component product of the CAPIOX EBS Circuit with X coating). If a centrifugal pump
other than the dedicated centrifugal pump is used, this may result in decreased performance.
Caution
• If there is physical damage, such as a crack, to the centrifugal pump, do not
use the centrifugal pump.
Caution
Attaching the Centrifugal Pump to the Drive Motor
• The Drive Motor contains magnets, and therefore do not bring metal objects
or magnets near the device. Malfunction may occur.
Warning
Check
Blood outlet port
Band
Centrifugal pump
Blood inlet port
• Prevent physical shock to the pump receptacles of the Drive Motor. The
internal magnets of the device may be damaged.
• When the centrifugal pump is not used, attach the magnet cover to the pump
receptacle of the Drive Motor.
Cautions
Slide hook
Magnet cover
Drive Motor
Fixed hook
Blood outlet
port
Rim
Fixed hook
Drive Motor
Centrifugal pump
Slide hook
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...