Various menu settings
4.
Tap [Alarm] of the lower limit. (Figure 3-4)
5. Tap
or
to input the lower limit value you want to set.
(Figure 3-5)
“
(red)” indicates the threshold value of the Low Flow Alarm in accordance
with an input flow value.
Figure 3-4
6.
Tap [OK] to complete the setting. (Figure 3-6)
Figure 3-5
Figure 3-6
• If you do not tap [OK], the changes to the setting will not take effect.
• The Low Flow Alarm is issued when flow has fallen below the set threshold value. When the lower limit value is set at 0.0 LPM
(L/min) and the flow is 0.00 LPM (L/min), the Low Flow Alarm is not issued.
The following settings are available:
• The on/off status of the pressure measuring function
• The threshold values of Pressure Alarm and Pressure Alert (setup range: -250 to 900 mmHg [-33.3 to 120.0 kPa])
1.
Tap the Pressure Alarm icon. (Figure 3-7)
The Pressure Alarm screen opens.
Figure 3-7
85
Notes
Setting Pressure Measuring Function