Precautions
Cautions
• When used in the vicinity of an electrosurgical knife (Medical radio knives are surgical equipment for incision and
coagulation by high energy radio frequency current), it may cause malfunction due to the noise of high-frequency current.
Check the following before use.
(1) Electrosurgical knives have a different level of high-frequency noise emission depending on the types. Do not use old
models (vacuum tube gap type) in combination particularly as the noise levels from those are higher.
(2) The distance from the electrosurgical knife cord (knife holder, knife cord and return electrode cord) and electrosurgical
knife body to this system should be kept as far away as possible (25 cm or more).
(3) The electrosurgical knife and this system should be operated by a power supply from a separate system, and both
should be securely grounded.
• When an electrosurgical knife is used, ensure a return electrode is properly attached. [If an electrosurgical knife is used
while a return electrode is not properly attached, burning of the electrode attaching part, a failure or malfunction of this
system may occur.]
(Controller)
• Make sure to use CDI System 500 (TCVS product codes: 500A, 500AV, 500AHCT, 500AVHCT) for the CDI port of the
Controller. [Otherwise an electric shock or short circuit may occur resulting in failure of this system.]
• Do not connect the USB port of the Controller to any device that is not bus powered. [Otherwise an electric shock or short
circuit may occur resulting in failure of this system.]
• Do not allow the blood circuit, etc. to contact the LCD touch panel display. [If the blood circuit, etc. mistakenly contacts the
screen, system malfunction may occur.]
• The air bubble detector function is intended to detect air bubbles entering the blood inlet line. Therefore when the air
bubble detector function turns on, attach the ABD/Flow Sensor at the blood inlet port of the centrifugal pump. [As air
bubbles entering the blood inlet line are destroyed and dispersed by the centrifugal pump, the air bubbles may not be
detected if the ABD/Flow Sensor is attached at the blood outlet line.]
• Set the alarm threshold and air bubble detection sensitivity and record the settings. In addition, if the alarm threshold is
set to be extremely large or extremely small, the alarm may not work. Be careful to it. [If the setting is wrong, the Controller
may not operate as intended.]
• The display screen and the Operation panel (buttons, etc.) of the Controller should not be pressed with excessive force
and should not be operated with a pointed object. [It may cause damage to or failure of the display screen or the Operation
panel.]
• Do not use this system if it is exposed to strong light such as direct sunlight. [This may obstruct you seeing of the LCD
touch panel screen.]
• When the Controller is transported while in use, do not touch the buttons, etc. inadvertently or use the key lock function.
[Otherwise it may result in an unintended operation (stop, start, power on/off).]
• Prior to first use of the Controller, or if unused for a long period, connect this system to an AC power supply and provide a
sufficient charge (6 hours or more). [If not sufficiently charged, this system may not be able to operate using the internal
battery during a power failure, etc.]
• When the Low Battery Alarm or the Battery Empty Alarm is issued, immediately connect the Controller to the AC power
source. [As this system may stop, health damages may occur on the patient.]
• When charging the Controller during storage, ensure the charge lamp in the control panel is lit. [The charge may not be
performed properly, which may result in no power supply in an emergency situation.]
• When charging the Controller during storage, charge it after turning the power off. [If the power is ON, it will continue to
update the history information and data will be deleted from the oldest case and cannot be confirmed.]
• When using the external communication function, pay particular attention, as it is more susceptible to the effect from an
electrosurgical knife, mobile phone, radio device, defibrillator, etc. Regularly check that this system is operating normally.
[The Controller may become vulnerable to the influence.]
• When the external communication function is used, pay attention to the operating hours on battery. [Operating hours will
be shortened.]
(Backup Controller)
• When moving the system while the Backup Controller is being used, do not touch switches carelessly. [An unintentional
operation (stop, start or ON/OFF of power) may occur.]
• Before pressing the power switch of the Backup Controller, confirm that the motor speed knob is set at “0”. [If the motor
speed knob is not set to “0”, the Drive Motor will immediately drive with that setting value.]
• Do not use any Drive Motor other than the included Drive Motor or the Drive Motor sold separately (Catalogue No.:
ME
*
SP200M) with the Backup Controller. [This may result in decreased performance. In addition, influence may be given
to the emission level of electromagnetic wave and the tolerance of this system.]
23
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...