Precautions
Precautions Regarding Installation on Ambulance
<Power consumption examples>
CAPIOX Centrifugal Pump Controller SP-200
Maximum 250VA (Steady-state)
Maximum 400VA (Peak: when the Drive Motor is rapidly accelerating, auto-priming is operating)
1. Ensure the battery of this system is fully charged.
• All segments of the battery symbols on both the display and control panel are fully lit.
2. After connecting this system to a power source in an ambulance, ensure that AC power is being supplied to this system.
• The AC power lamp is lit.
When the AC power lamp is turned off, the system is operating on battery. (Refer to page 72.)
3. Ensure the Hand Crank is prepared.
26
• When installing the system in an ambulance and transferring a patient in the event of an emergency, make sure the power
supply of the ambulance is always higher than the total consumption power for all electric devices installed. [If the supply
power of the ambulance becomes insufficient, the device shuts down and circulation cannot be maintained.]
Warning
• Avoid sudden temperature changes even when operating within the system
ʼ
s indicated temperature specifications.
[Condensation inside the system results in damage, time degradation, and decreased performance.]
Caution
Warnings
• Always prepare a backup system (the Backup Controller, the Hand Crank, etc.) to maintain circulation during an
emergency. [Circulation cannot be maintained during an emergency.]
• Monitor the circulation status (the operation status of the system, blood storage state of a reservoir, etc.) during
extracorporeal circulation without fail. [1. As this system does not have a function to monitor the status of the patient, the
system cannot detect when the status of the patient changes even if the system operates properly. 2. The system cannot
detect the leakage of liquid (including blood) caused by damage to the extracorporeal circulation circuit. 3. It is possible
that the system will not detect a flow abnormality, even if the ABD/Flow Sensor is used. 4. This system may not operate in
accordance with the specifications if the system is used together with another system.]
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...