To ensure safe operation and the longest possible life, perform periodic maintenance and inspections. If any fault is detected,
immediately stop using the system and contact Terumo’s service representative.
Maintenance and Inspections by Users
This section describes the inspection items, inspection method and inspection cycle.
Prepare the following before inspection:
1.
Vaseline (used in the inspection before circulation starts)
2.
Stopwatch (used in the six-month inspection)
Confirm the contents in the following sequence before each use:
• Inspection of the external appearance
• Inspection of the start-up sequence
• Inspection of the battery-powered operation function
• Inspection of the Drive Motor Disconnected Alarm function and the ABD/Flow Sensor Disconnected Alarm function
Inspection method of the external appearance (Before each use)
1.
There are no significant scratches or damage in the appearance.
2.
There are no scratches or pinch marks on the outer covering of the cables.
3.
There are no foreign items or damage in the connectors.
4.
Turn the motor speed knob and check that it turns smoothly.
122
Maintenance and Inspections
• Since this system is precision equipment, it should not be used if it has received any impact (drop to floor, falling, violent
shock). Even though no fault is observed in the system appearance, internal components may damage. This may result
in decreased performance (flow rate accuracy and various alarm functions, etc.), and therefore inspection is required.
Contact Terumo’s service representative for inspection.
• Do not disassemble, make alterations (including actions that interfere with the functionality or performance such as taping
the LCD or a movable part) to, or repair this system. This may result in failure, damage or device performance degradation
of this system.
• Execute maintenance and inspections periodically for devices in the backup kit. When executing maintenance and
inspections, use the Drive Motor and the AC power cable included in the backup kit. Due to a low frequency of use, the
abnormality may not be detected.
• Check displays to ensure this system will operate properly.
Cautions
Items to Prepare
Inspection Before Each Use
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...