Operation Procedure
3.
Attach the CAPIOX EBS Circuit Holder to the IV pole of the
Cart and secure it with the knob. (Figure 1-9)
• The CAPIOX EBS Circuit Holder is secured.
Figure 1-9
4.
Remove the plug cap of the Drive Motor. Insert the
plug into the drive motor connector on the rear panel.
(Figure 1-10
a
) Then, turn the lock ring clockwise until you
hear a click. (Figure 1-10
b
)
When removing the plug of the Drive Motor, turn the lock ring counter-
clockwise until you hear a click, and pull it out straight.
Figure 1-10
Using the pole clamp that comes with the Drive Motor
1.
Attach the pole clamp to the IV pole of the Cart and secure
it with the knob. (Figure 1-11
a
)
• The pole clamp is secured.
2.
Adjust the angle of the motor with the angle adjustment
knob. (Figure 1-11
b
)
Figure 1-11
34
Check
Check
IV pole
Knob
CAPIOX EBS Circuit Holder
Drive motor connector
Lock ring
Plug
Plug cap
Tighten
IV pole
Loosen
Knob
Pole clamp
Angle adjustment knob
Slide hook