Maintenance and Inspections
Serial number:
Date:
/ /
Person in charge:
Check items
Inspection before circulation starts (Before each use)
Inspection of the
motor drive (Before
each use)
• Are there any abnormalities such as liquid leakage or rotation deflection of the
centrifugal pump or abnormal sounds from the Drive Motor while the blood outlet line
of the centrifugal pump is clamped and the centrifugal pump is rotated at a speed of
3000 RPM for approximately 30 seconds?
• Does the centrifugal pump stop when the motor speed is changed to 0 RPM?
Pass
Pass
Fail
Fail
Inspection of the ABD/
Flow Sensor (Before
each use)
• Is the flow displayed in the flow indicator in the LCD when the centrifugal pump is
rotated?
Pass
Fail
Inspection of the Fast
Clamp (Before each
use)
• Is the plunger closed (up position) after the Fast Clamp is operated by tapping the
[Clamp]?
• Does the lever return to the original position after pulling it up until you hear a click? If
not, does the lever return to the original position when it is pulled down?
Pass
Pass
Fail
Fail
Overall judgement
(If any one item is judged as fail, repairs are necessary. Immediately stop using the system and contact
Terumo’s service representative.)
Pass
Fail
Remarks:
Serial number:
Date:
/ /
Person in charge:
Check items
Six-month inspection (Once every six months)
Inspection of the
battery (Once every
six months)
• Is there more than 70% battery charge level when the battery is charged until the
charge lamp turns off?
• Can the system operate for more than three hours on battery power with the motor
speed set at 3000 RPM?
• Does the charge lamp turn off within six hours when charging starts after the power is
turned off?
Pass
Pass
Pass
Fail
Fail
Fail
Overall judgement
(If any one item is judged as fail, repairs are necessary. Immediately stop using the system and contact
Terumo’s service representative.)
Pass
Fail
Remarks:
130
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...