Operation Procedure
Cross-
section
diagram
Air bubble
Direction of
gravitational
force
Sensor
detection
direction
Direction of circulation
Tube
Detection window
of the ABD/Flow Sensor
Cross-
section
diagram
Sensor
detection
direction
Air bubble
Direction of
gravitational
force
2.
Attach the clamp-on ABD/Flow sensor to the blood circuit
tube. (Figure 1-35, Figure 1-36)
Fully insert the blood circuit tube into the detection window of the clamp-on
ABD/Flow sensor.
Figure 1-35
CORRECT
WRONG
Figure 1-36
45
Clamp-on ABD/Flow sensor
Tube
Direction of
circulation
• When the air bubble detection sensitivity is high, attach the sensor so that
the ABD/Flow Sensor detection window and the direction of gravitational
force are orthogonal. It may not correctly detect air bubbles. (Figure 1-36)
Caution