Remote Control Commands
R&S
®
SMF100A
523
Operating Manual 1167.2319.02 ─ 12
[:SOURce<hw>]:POWer:ALC:SOURce
............................................................................ 524
[:SOURce<hw>]:POWer:ALC[:STATe]
............................................................................. 524
[:SOURce<hw>]:POWer:ATTenuation:RFOFf:MODE
.........................................................525
......................................................................................525
...............................................................................525
[:SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude]
..................................................526
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet
........................................................526
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:RCL
.............................................................527
[:SOURce<hw>]:POWer:LIMit[:AMPLitude]
.......................................................................527
.....................................................................................528
.......................................................................................529
......................................................................................529
[:SOURce<hw>]:POWer:RESolution
................................................................................ 529
[:SOURce<hw>]:POWer:SPC:CRANge
............................................................................ 530
............................................................................... 530
................................................................................ 530
[:SOURce<hw>]:POWer:SPC:SELect
.............................................................................. 531
...............................................................................531
[:SOURce<hw>]:POWer:SPC:TARGet
............................................................................. 531
.......................................................................................532
[:SOURce<hw>]:POWer:STEP[:INCRement]
.................................................................... 532
............................................................................. 533
........................................................................................533
[:SOURce<hw>]:POWer:ALC:PMETer:TYPE
<Type>
The command selects the power meter used for external level control.
Parameters:
<Type>
NRP | NRVD | HP436 | HP437 | HP438 | ML2438 | P44X
*RST:
NRP
Example:
POW:ALC:PMET:TYPE HP438
the specified power meter is used for external level control.
Manual operation:
See
" Power Meter Type - External Level Control "
[:SOURce<hw>]:POWer:ALC:REFerence
<Reference>
The command sets the reference voltage for external power control. See specification
of the used instrument for external power control for details.
Parameters:
<Reference>
float
Range:
0 to 2.5
Increment: 1E-4
*RST:
1.5
Example:
SOUR:POW:ALC:REF 0.37
sets a refebrence voltage of 0.37 volt.
Manual operation:
See
" Reference Voltage - External Level Control "
SOURce Subsystem