Remote Control Commands
R&S
®
SMF100A
477
Operating Manual 1167.2319.02 ─ 12
[:SOURce]:INPut:TRIGger:SLOPe
<Slope>
Sets the polarity of the active slope of an externally applied trigger signal at the trigger
input (BNC connector at the rear of the instrument).
The setting is effective for both inputs at the same time.
Parameters:
<Slope>
NEGative | POSitive
*RST:
POSitive
Example:
INP:TRIG:SLOP NEG
Activates the falling slope of the external trigger signal at the
trigger input.
Manual operation:
See
7.13.8 SOURce:LFOutput Subsystem
The
SOURce:LFOutput
subsystem contains the commands for setting the LF signal
source in CW and Sweep mode and for analog modulation.
Two internal LF generators and a noise generator are available with option R&S
SMF-
B20.
The suffix for
LFOutput<ch>
denotes the selected LF generator. The source for the
LF sweep (LF generator 1 or 2) is selected with command
SOURce:LFOutput:SWEep:FREQuency:SOURce
.
Example
The following example shows how to set an LF sweep.
1. Set the sweep range.
LFOutput:FREQuency:STARt 4 kHz
LFOutput:FREQuency:STOP 10 kHz
2. Select linear or logarithmic sweep spacing.
LFOutput:SWEep[:FREQuency]:SPACing LIN
3. Set the step width and dwell time.
LFOutput:SWEep[:FREQuency]:STEP[:LINear] 100 Hz
LFOutput:SWEep[:FREQuency]:DWELl 20 ms
4. Determine the sweep mode.
LFOutput:SWEep:MODE AUTO
5. Determine the trigger.
TRIGger0:SOURce SINGle
6. Activate the sweep.
LFOutput:FREQuency:MODE SWEep
7. Trigger the sweep (depending on the mode).
SOURce Subsystem