Remote Control Commands
R&S
®
SMF100A
475
Operating Manual 1167.2319.02 ─ 12
Example:
FREQ:STOP 2 GHz
sets the stop frequency for the frequency sweep to 2 GHz.
FREQ:STAR 1 MHz
sets the start frequency for the frequency sweep to 1 MHz.
Manual operation:
See
" Stop Freq - Frequency Sweep "
[:SOURce<hw>]:FREQuency:STEP[:INCRement]
<Increment>
Sets the step width for
To adjust the frequency step by step with this step size, use the
FREQ:UP
and
FREQ:DOWN
commands.
Note:
This value also applies to the step width of the rotary knob of the instrument and
increases or decreases the frequency accordingly, when you work in user-defined step
mode.
Parameters:
<Increment>
float
Range:
full frequency range
Increment: see the data sheet: RF characteristics > Resolution
of setting
*RST:
1E6
Example:
FREQ:STEP 50 kHz
sets the step width for the frequency setting to 50 kHz.
Manual operation:
See
[:SOURce<hw>]:FREQuency:STEP:MODE
<Mode>
Activates (
USER
) or deactivates (
DECimal
) the user-defined step width used when
varying the frequency value with the frequency values
UP/DOWN
. The command is
linked to the command "Variation Active" for manual control, i.e. the command also
activates/deactivates the user-defined step width used when varying the frequency
value with the rotary knob.
Parameters:
<Mode>
DECimal | USER
*RST:
DECimal
Example:
FREQ:STEP 50 kHz
sets the step width for the frequency setting to 50 kHz.
FREQ:STEP:MODE USER
actives this step width for frequency variation with the rotary
knob (manual control) and with frequency values
UP/DOWN
(remote control).
Manual operation:
See
SOURce Subsystem