Remote Control Commands
R&S
®
SMF100A
526
Operating Manual 1167.2319.02 ─ 12
Manual operation:
See
" Display Level as Voltage of EMF - RF Level "
[:SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude]
<Amplitude>
Sets the RF level applied to the DUT.
Notes:
If specified, a level offset
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:
is included according to the formula:
Minimum level + OFFSet ... Maximum level + OFFSet
In addition to numerical values, you can increase or decrease the values step by step
with the
UP
and
DOWN
according to the step width defined with
.
The RF output is activated with
OFF).
Parameters:
<Amplitude>
Minimum level ... Maximum level
Determines the RF output level.
Range:
Minimum level to Maximum level
*RST:
-30
Example:
The keywords of this command are largely optional. Therefore,
both the long and short form of the command are shown.
SOUR:POW:LEV:IMM:AMPL 15
or
:POW 15
sets the RF level at output A to 15 dBm.
Manual operation:
See
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet
<Offset>
Note:
The level offset is also effective for level sweeps!
Specifies the constant level offset of a downstream attenuator/amplifier. If a level offset
is entered, the level entered with
:POWer
no longer corresponds to the RF output
level.
The following correlation applies:
POWer
= RF output level +
POWer:OFFSet
.
Entering a level offset does not change the RF output level, but rather the query value
of
:POWer
.
For more information, see
Chapter 5.4.1.1, "RF level vs. RF output level"
Only dB is permitted as the unit here. The linear units (V, W, etc.) are not permitted.
SOURce Subsystem