![Rohde & Schwarz SMF-B1 Operating Manual Download Page 481](http://html1.mh-extra.com/html/rohde-and-schwarz/smf-b1/smf-b1_operating-manual_1477740481.webp)
Remote Control Commands
R&S
®
SMF100A
466
Operating Manual 1167.2319.02 ─ 12
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
SOUR:DM:PSK:STAT ON
activates the phase modulation.
Options:
R&S SMF-B20
Manual operation:
See
7.13.5 SOURce:FM Subsystem
The FM subsystem contains the commands for checking the frequency modulation.
Characteristics which are valid for all modulations and the LF Output are configured in
the
SOURce:LFOutput
subsystem (e.g. frequency). The external signal is input at the
EXT1 and/or EXT2 connector. Internal and external modulation source can be selected
at the same time, thus enabling two-tone FM modulation.
For information about the required options, see
Chapter 5.5.3, "Frequency Modulation
[:SOURce<hw>]:FM<ch>[:DEViation]
............................................................................... 466
.............................................................................................467
............................................................................................. 467
.................................................................................. 467
..................................................................................... 468
[:SOURce<hw>]:FM<ch>[:DEViation]
<Deviation>
Sets the deviation of the frequency modulation signal in Hz. The maximum deviation
depends on the set RF frequency and the selected modulation mode (see data sheet).
Suffix:
<ch>
.
1..2
determines the modulation signal channel.
Parameters:
<Deviation>
float
Range:
see data sheet
Increment: 0.1
*RST:
10E3
Example:
FM1:DEV 3kHz
sets 3 kHz deviation for the frequency modulation signal in chan-
nel 1 (FM1).
Options:
R&S
SMF-B20
Manual operation:
See
SOURce Subsystem