Remote Control Commands
R&S
®
SMF100A
561
Operating Manual 1167.2319.02 ─ 12
Example:
TRIG0:SWE:SOUR SING
FREQ:MODE SWE
SWE:SHAP SAWT
SWE:RETR ON
activates retrace function, i.e. the frequency changes to the
value at start frequency while waiting for the next trigger event.
Manual operation:
See
" Retrace - RF Frequency Sweep "
[:SOURce<hw>]:SWEep[:FREQuency]:RUNNing?
Queries the current state of the frequency sweep mode.
Return values:
<State>
0 | 1 | OFF | ON
Example:
SWE:RUNN?
Response
"1"
: signal generation in level sweep active.
Usage:
Query only
[:SOURce<hw>]:SWEep[:FREQuency]:SHAPe
<Shape>
Sets the cycle mode for a sweep sequence (shape).
Parameters:
<Shape>
SAWTooth | TRIangle
SAWTooth
One sweep runs from start to stop frequency. Each subsequent
sweep starts at the start frequency, i.e. the shape of the sweep
sequence resembles a sawtooth.
TRIangle
One sweep runs from start to stop frequency and back, i.e. the
shape of the sweep resembles a triangle. Each subsequent
sweep starts at the start frequency.
*RST:
SAWTooth
Example:
SOUR:SWE:SHAP TRI
selects the sweep cycle with alternating ascending and
descending sweep directions.
Manual operation:
See
" Shape - RF Frequency Sweep "
[:SOURce<hw>]:SWEep[:FREQuency]:SPACing
<Spacing>
Selects the mode for the calculation of the frequency sweep intervals. The frequency
increases or decreases by this value at each step.
The keyword
[:FREQuency]
can be omitted. Then the command is SCPI-compliant.
SOURce Subsystem