Remote Control Commands
R&S
®
SMF100A
565
Operating Manual 1167.2319.02 ─ 12
[:SOURce<hw>]:SWEep:POWer:DWELl
<Dwell>
Sets the time taken for each level step of the sweep.
Tip:
It is recommended to switch off the "Display Update" for optimum sweep perform-
ance especially with short dwell times (
Parameters:
<Dwell>
float
Range:
1E-3 to 100
Increment: 100E-6
*RST:
15E-3
Example:
SWE:POW:DWEL 12 ms
sets a dwell time of 12 ms for a level sweep at the RF output.
Manual operation:
See
[:SOURce<hw>]:SWEep:POWer:EXECute
Triggers a sweep.
The sweep to be executed depends on the set sweep mode (
SOURce:SWEep:POWer:
MODE
). The command corresponds to the manual-control command "Execute Trigger".
Example:
SWE:POW:MODE STEP
sets the STEP trigger mode, i.e. a trigger starts the sweep ini-
tially, and then the sweep is generated continuously.
SWE:POW:EXEC
triggers a level sweep at the RF output.
Usage:
Event
Manual operation:
See
" Execute Single Sweep - Level Sweep "
[:SOURce<hw>]:SWEep:POWer:MARKer:ACTive
<Active>
The command activates the selected marker. The active marker is output with an
higher voltage than all other markers.
Parameters:
<Active>
NONE | M01 | M02 | M03 | M04 | M05 | M06 | M07 | M08 | M09 |
M10
*RST:
NONE
Example:
SWE:POW:MARK:ACT M05
sets marker 5 to be the active marker.
Manual operation:
See
" Active Marker - RF Level Marker "
SOURce Subsystem