Instrument Function
R&S
®
SMF100A
268
Operating Manual 1167.2319.02 ─ 12
In the upper section of the menu, the modulation source is selected and the modulation
switched on. The configuration of the selected external and/or internal modulation
source is performed in the lower section of the menu.
The menu varies according to the selected modulation source.
In case of modulation source pulse generator, the settings for the pulse characteristics
and the trigger are provided in the "Pulse Generator" section of the "Pulse Modulation"
dialog. Option R&S
SMF-K23 provides extended features, for example generation of
double pulse or selection of trigger mode. For description of the pulse generator dialog,
see
Chapter 5.6.3, "Pulse Generator"
Option R&S
SMF-K27 enables the generation of pulse trains. For description of the
Chapter 5.6.3.2, "Pulse Train Generation"
In the "Input/Trigger Settings" section, the external input and trigger are configured.
State
Activates pulse modulation.
If modulation source Pulse Generator is selected (requires option R&S
SMF-K23), the
pulse generator is switched on/off, in addition.
When the internal modulation source (pulse generator) is selected, the pulse generator
is switched on automatically and the video/sync signal is output at the PULSE VIDEO
output at the rear of the instrument. Signal output can be switched off in the "Pulse
Generator" dialog (see
Chapter 5.6.3, "Pulse Generator"
Remote command:
Source
Selects the modulation signal source for pulse modulation.
Modulation