Remote Control Commands
R&S
®
SMF100A
562
Operating Manual 1167.2319.02 ─ 12
Parameters:
<Spacing>
LINear | LOGarithmic | RAMP
LINear
With the linear sweep, the step width is a fixed frequency value
which is added to the current frequency. The step width for lin-
ear sweep is entered in Hz (see
LOGarithmic
With the logarithmic sweep, the step width is a constant fraction
of the current frequency. This fraction is added to the current fre-
quency. The logarithmic step width is entered in % (see
SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic
RAMP
With the ramp sweep, a synthesized continuous analog fre-
quency sweep is provided for a set sweep time. The sweep time
is entered under
[:SOURce<hw>]:SWEep[:FREQuency]:
*RST:
LINear
Example:
SWE:SPAC LIN
selects linear sweep spacing for a frequency sweep at the RF
output.
Manual operation:
See
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear]
<Linear>
Sets the step size for linear RF frequency sweep steps.
This parameter is related to the number of steps (
) within the sweep range as follows:
f
STARt
< f
STOP
freq_points = (f
SPAN
/ step_lin) + 1
If you change the step size, the number of steps changes accordingly. The sweep
range remains the same.
The keywords
[:FREQuency]
and
[:LINear]
can be omitted. The command is then
SCPI-compliant.
Parameters:
<Linear>
float
Range:
full frequency range
Increment: see the data sheet: RF characteristics > Resolution
of setting
SOURce Subsystem