Remote Control Commands
R&S
®
SMF100A
564
Operating Manual 1167.2319.02 ─ 12
[:SOURce<hw>]:SWEep[:FREQuency]:TIME
<Time>
Sets the sweep time for ramp sweep.
Parameters:
<Time>
float
Range:
0.01 to 100
Increment: 1E-4
*RST:
0.015
Example:
SOUR:FREQ:TIME 10ms
sets a sweep time of 10 ms for a ramp sweep.
Options:
R&S SMF-K4
Manual operation:
See
" Sweep Time - RF Frequency Sweep "
[:SOURce<hw>]:SWEep:MARKer:OUTPut:POLarity
<Polarity>
The command selects the polarity of the marker signal.
Parameters:
<Polarity>
NORMal | INVerted
NORMal
The marker level switches to HIGH after reaching the mark.
INVerted
The marker level switches to LOW after reaching the mark.
*RST:
NORMal
Example:
SWE:FREQ:MARK:OUTP:POL NORM
sets the marker output polarity to normal.
Options:
R&S SMF-B20
Manual operation:
See
[:SOURce<hw>]:SWEep:POWer:AMODe
<AMode>
Selects the ranges of level settings for the level sweep. The sweep is either performed
in the low level or in the high level ranges.
Parameters:
<AMode>
AUTO | FIXed | NORMal | HPOWer | MANual
NORMal
The level settings are made in the range of the electronically
switching attenuator. The high level ranges are not available.
HPOWer
The level settings are made in the high level range.
*RST:
NORMal(HighPower)|AUTO
Example:
SWE:POW:AMOD HPOW
selects the high level ranges for level sweep.
SOURce Subsystem