Instrument Function
R&S
®
SMF100A
291
Operating Manual 1167.2319.02 ─ 12
"Extern Start/
Stop"
Sets an automatically repeated sweep cycle that is started, stopped
and restartet by subsequent external trigger events.
The first external trigger signal starts the sweep (Start).
The next external trigger signal stops the sweep at the current LF fre-
quency (Stop).
The third external trigger signal starts the sweep at the start fre-
quency (Start).
Refer to the description of the rear panel for information about the
connectors for external trigger signal input (see
Example:
SOUR:LFO:SWE:VOLT:MODE AUTO
TRIG:LFVS:SOUR EAUT
SOUR:LFO:VOLT:MODE SWE
(External trigger)
Remote command:
[:SOURce]:LFOutput:SWEep:VOLTage:MODE
[:SOURce]:LFOutput:VOLTage:MODE
Specific Sweep Parameters
The sweep parameters described in this topic are only available for particular emula-
tion languages.
For details, see
"Remote Emulation with the R&S
SMF microwave signal gen-
erator".
Mode ← Specific Sweep Parameters
Selects the RF frequency sweep mode.
("Auto" = timer controlled, "Manual" = manually controlled)
Repeat ← Specific Sweep Parameters
Sets the repetition mode.
("On" = continuously; "Off" = once)
Start Trigger ← Specific Sweep Parameters
Selects the mode to initiate a sweep.
("Auto" = starts immediately; "Single" = starts with a command (Bus); "External" =
starts with an external trigger signal)
Point Trigger ← Specific Sweep Parameters
Selects the mode to trigger the progress of each step.
("Auto" = immediately; "Single" = remotely (Bus); "External" = external signal)
Execute Single Sweep - LF Level Sweep
Triggers the sweep manually. A manual sweep can only be triggered if "Mode Single"
is selected.
Modulation Generator, Pulse Generator and LF Output