![Rohde & Schwarz SMF-B1 Operating Manual Download Page 253](http://html1.mh-extra.com/html/rohde-and-schwarz/smf-b1/smf-b1_operating-manual_1477740253.webp)
Instrument Function
R&S
®
SMF100A
238
Operating Manual 1167.2319.02 ─ 12
"Distal / Mesial / Proximal Power"
Display the absolute power values of the medial, low and high refer-
ence level in dBm.
Remote command:
:TRACe<ch>[:POWer]:SWEep:MEASurement:POWer:LREFerence:DISPlay:
ANNotation[:STATe]
:TRACe<ch>[:POWer]:SWEep:MEASurement:POWer:LREFerence?
:TRACe<ch>[:POWer]:SWEep:MEASurement:POWer:HREFerence:DISPlay:
ANNotation[:STATe]
:TRACe<ch>[:POWer]:SWEep:MEASurement:POWer:HREFerence?
Diagram.. - Power Analysis
Returns to the "NRP-Z Analysis Diagram".
Configure Trigger
The "Trigger…" button opens the "Trigger NRP-Z Analysis" dialog. The button is only
active for time measurement mode. For this mode, the measurement start has to be
known to the sensor as the measurement is controlled by the sensor.
Power Sensor - Power Analysis
Selects the power sensor to be set if more than one sensor is connected to the instru-
ment.
Remote command:
n.a.
In remote control mode, the sensor is selected via the numeric suffix in the sense key
word of the command, for example
SENSe2:POWer:SWEep:…
.
Mode - Power Analysis
Selects if the measurement is free running, or starts only after an internal or external
trigger event.
Remote command:
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:TRIGger:SOURce
Level - Power Analysis
Sets the trigger threshold. This setting is also possible by means of the trigger marker
on the left side of the diagram.
RF Level