Remote Control Commands
R&S
®
SMF100A
415
Operating Manual 1167.2319.02 ─ 12
:SENSe[:POWer]:SWEep:HCOPy:FILE[:NAME]:AUTO[:FILE]:DAY:STATe
<State>
Activates the usage of the day in the automatic file name.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
1
Example:
SENS:SWE:HCOP:FILE:AUTO:DAY:STAT OFF
deactivates the usage of the day in the automatic file name.
Options:
Option R&S SMF-K28
Manual operation:
See
" Save Options - Power Analysis "
:SENSe[:POWer]:SWEep:HCOPy:FILE[:NAME]:AUTO[:FILE]:MONTh?
Queries the day of the date part in the automatic file name.
Return values:
<Month>
integer
Range:
1 to 12
*RST:
1
Example:
SENS:SWE:HCOP:FILE:AUTO:MONT?
queries the month of the date part in the automatic file name.
Usage:
Query only
Options:
Option R&S SMF-K28
Manual operation:
See
" Save Options - Power Analysis "
:SENSe[:POWer]:SWEep:HCOPy:FILE[:NAME]:AUTO[:FILE]:MONTh:STATe
<State>
Activates the usage of the month in the automatic file name.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
1
Example:
SENS:SWE:HCOP:FILE:AUTO:MONT:STAT OFF
deactivates the usage of the month in the automatic file name.
Options:
Option R&S SMF-K28
Manual operation:
See
" Save Options - Power Analysis "
:SENSe[:POWer]:SWEep:HCOPy:FILE[:NAME]:AUTO[:FILE]:NUMBer?
Queries the generated number in the automatic file name.
Power Sensor Measurement Subsystems