Remote Control Commands
R&S
®
SMF100A
584
Operating Manual 1167.2319.02 ─ 12
:SYSTem:COMMunicate:SERial:PARity
<Parity>
Sets the parity for the serial remote control interface.
Parameters:
<Parity>
NONE | ODD | EVEN
*RST:
NONE
Example:
SYST:COMM:SER:PAR NONE
selects parity NONE.
Manual operation:
See
:SYSTem:COMMunicate:SERial:SBITs
<SBits>
Sets the number of stop bits for the serial remote control interface.
Parameters:
<SBits>
1 | 2
*RST:
1
Example:
SYST:COMM:SER:SBIT 2
selects 2 stop bits.
Manual operation:
See
:SYSTem:COMMunicate:SOCKet:RESource?
Queries the visa resource string for remote control via LAN interface, using TCP/IP
socket protocol.
Return values:
<Resource>
string
Example:
SYSTem:COMMunicate:SOCKet:RESource?
Response:
"TCPIP::10.113.1.150::5025::SOCKET"
Usage:
Query only
Manual operation:
See
:SYSTem:IDENtification
<Identification>
Selects the mode the instrument identification is performed.
Parameters:
<Identification>
AUTO | USER
AUTO
The "IDN String" and the "OPT String" are set automatically.
USER
Enables the selection of user definable "IDN String" and "OPT
String".
*RST:
AUTO
SYSTem Subsystem