Remote Control Commands
R&S
®
SMF100A
502
Operating Manual 1167.2319.02 ─ 12
Parameters:
<Manual>
float
Minimum level ... Maximum level
The value range for the level setting varies according to the
instrument model. The values are given in the data sheet.
Range:
0 to 6
Increment: 0.001
*RST:
1
Example:
SOUR:LFO:SWE:VOLT:MODE MAN
sets the Step sweep mode.
SOUR:LFO:VOLT:MAN -5dBm
sets an RF level of -5 dBm for the next setting in the Step sweep
mode.
SOUR:LFO:VOLT:MODE SWE
sets the Level Sweep mode.
SOUR:LFO:VOLT:MAN -5.5dBm
triggers the next sweep step with a level of -5.5 dBm.
Options:
R&S SMF-B20
Manual operation:
See
" Current Level - LF Level Sweep "
[:SOURce]:LFOutput:VOLTage:MODE
<Mode>
The command sets the instrument operating mode and therefore also the commands
used to set the LF output level.
Parameters:
<Mode>
CW | FIXed | SWEep
CW
The instrument operates at a constant level. The output level is
set with
:SOURce:LFOutput:POWer
.
SWEep
The instrument operates in SWEep mode. The level is set using
the commands
SOURce:LFOutput:POWer:STARt; STOP;
MANual
.
*RST:
CW
Example:
SOUR:LFO:VOLT:MODE SWE
sets the SWEep mode. The settings under
SOURce:LFOutput:POW:STARt; STOP; MANual
become
effective.
Options:
R&S SMF-B20
Manual operation:
See
[:SOURce]:LFOutput:VOLTage:STARt
<Start>
Sets the start level value for the LF sweep.
SOURce Subsystem