Remote Control Commands
R&S
®
SMF100A
497
Operating Manual 1167.2319.02 ─ 12
Manual operation:
See
[:SOURce<hw>]:LFOutput<ch>:SHAPe
<Shape>
Selects the shape of the LF signal.
Parameters:
<Shape>
SINE | SQUare | TRIangle | TRAPeze
*RST:
SINE
Example:
LFO:SHAP SQU
selects a rectangular shape for the signal of the LF generator.
Options:
R&S SMF-B20
Manual operation:
See
[:SOURce]:LFOutput:SWEep:VOLTage:SOURce
<Source>
The command sets the trigger source for the LF level sweep.
The names of the parameters correspond directly to the various settings under manual
control. SCPI uses other names for the parameters; these names are also accepted by
the instrument. The SCPI names should be used if compatibility is an important consid-
eration.
An overview of the various names is given in the
Parameters:
<Source>
SINGle | BUS | AUTO | IMMediate | EXTernal | EAUTo
AUTO | IMMediate
The trigger is free-running, i.e. the trigger condition is fulfilled
continuously. As soon as one sweep is finished, the next sweep
is started.
SINGle | BUS
One complete sweep cycle is triggered by the GPIB bus com-
mands
[:SOURce]:LFOutput:SWEep:VOLTage:EXECute
,
the
*TRG
or
:TRIGger<hw>[:SWEep][:IMMediate]
on page 597. The mode has to be set to "AUTO"
(
SOUR:LFO:SWE:VOLT:MODE AUTO
).
EXTernal
The sweep is triggered externally via the TRIGGER connector or
the AUX connector at the rear of the instrument.
EAUTo
The sweep is triggered externally via the INST TRIG connector.
As soon as one sweep is finished, the next sweep is started. A
second trigger event stops the sweep at the current frequency, a
third trigger event starts the trigger at the start frequency, and so
on.
*RST:
AUTO
SOURce Subsystem