Remote Control Commands
R&S
®
SMF100A
472
Operating Manual 1167.2319.02 ─ 12
Parameters:
<Mode>
CW | FIXed | SWEep | LIST
CW|FIXed
Sets the fixed frequency mode.
CW
and
FIXed
are synonyms. The instrument operates at a
defined frequency, set with command
SWEep
Sets the sweep mode. The instrument processes the frequency
settings in defined sweep steps. To determine the corresponding
frequency values, use the commands
,
[:SOURce<hw>]:FREQuency:CENTer
and
LIST
Sets the list mode. The instrument processes the frequency and
level settings by means of values loaded from a list.
To configure the list mode settings use the commands of the
*RST:
CW
Example:
FREQ:MODE SWE
sets the SWEep mode.
Manual operation:
See
[:SOURce<hw>]:FREQuency:MULTiplier
<Multiplier>
Sets the value for the multiplication factor of a subsequent downstream instrument.
Parameters:
<Multiplier>
float
Range:
1 to dynamic
Increment: 0.001
*RST:
1
Example:
FREQ:MULT 1
sets the multiplication factor to 1.
Manual operation:
See
[:SOURce<hw>]:FREQuency:OFFSet
<Offset>
Sets the frequency offset of a downstream instrument, for example a mixer.
If you have specified an
OFFSet
and / or a
MULTiplier
factor, the actual frequency
at the RF output does not change, but rather the value queried with
according to the following formula:
f
FREQ
= f
RFout
* f
MULTiplier
+ f
OFFSet
SOURce Subsystem