Remote Control Commands
R&S
®
SMF100A
356
Operating Manual 1167.2319.02 ─ 12
Example:
SYST:FPR
All instrument settings (also the settings that are not currently
active) are reset to the factory values.
Usage:
Event
Manual operation:
See
7.4 CALibration Subsystem
The
CALibration
system contains the commands for adjustment. Adjustment is trig-
gered by the query commands. The response "0" indicates error-free adjustment, and
the response "1" means that an error occurred during adjustment.
:CALibration<hw>:ALL[:MEASure]?
................................................................................. 356
:CALibration<hw>:FREQuency[:MEASure]?
......................................................................356
:CALibration<hw>:LEVel:LOOPgain:STATe
...................................................................... 357
:CALibration<hw>:LEVel[:MEASure]?
...............................................................................357
:CALibration<hw>:LEVel:EXTern:DATA
............................................................................357
...................................................................................... 358
:CALibration:LFOutput[:MEASure]?
..................................................................................358
:CALibration<hw>:ROSCillator[:DATA]
............................................................................. 358
:CALibration<hw>:ALL[:MEASure]?
[<Force>]
Starts all internal adjustments that do not require external measurement equipment.
Query parameters:
<Force>
string
Return values:
<Measure>
select
Example:
CAL:ALL:MEAS?
// Response "0"
// Adjustment has been performed successfully
Usage:
Query only
Manual operation:
See
:CALibration<hw>:FREQuency[:MEASure]?
Starts all adjustments which affect the frequency.
Return values:
<Measure>
0 | 1
Example:
CAL:FREQ:MEAS?
starts the adjustments for maximum frequency accuracy.
Response: "0"
the adjustments have been performed successfully.
CALibration Subsystem