Instrument Function
R&S
®
SMF100A
103
Operating Manual 1167.2319.02 ─ 12
Adjust All
Performs all available internal calibration routines of the instrument.
Remote command:
:CALibration<hw>:ALL[:MEASure]?
Adjust Synthesis
Performs all adjustments which affect the frequency.
This includes adjustment of option R&S SMF-B20, AM/FM/SCAN modulator, LF and
Low Phase Noise.
Remote command:
:CALibration<hw>:FREQuency[:MEASure]?
Adjust Level
Performs all adjustments which affect the level. The acquired correction values
improve the settling time and the signal quality.
Remote command:
:CALibration<hw>:LEVel[:MEASure]?
Adjust Modulation LF Generator
Performs all adjustments that affect the internal modulation generators.
Remote command:
:CALibration:LFOutput[:MEASure]?
Use ext. Level Correction Data
Switch on or off use of external level correction data. This switch can be used in case
of corrupt external level correction data.
By default the instrument uses correction data obtained in the factory before delivery.
In addition, customer data can be used for external level correction. The customer data
is obtained using a R&S NRP power sensor. External level correction is a protected
function (see service manual, chapter 2, "Adjustment").
Remote command:
Use ext. Loop Gain Correction Data
Switch on or off use of external loop gain correction data.
This feature is a protected function ( see service manual, chapter "Adjustment").
Remote command:
:CALibration<hw>:LEVel:LOOPgain:STATe
Continue Adjustment On Error
Continues the internal adjustment automatically after an error was detected. By default,
internal adjustment stops on error (see service manual, chapter "Adjustment").
This function is a protected function and can be accessed if protection level 1 is dis-
abled.
Remote command:
n.a.
General Instrument Settings