Remote Control Commands
R&S
®
SMF100A
357
Operating Manual 1167.2319.02 ─ 12
Usage:
Query only
Manual operation:
See
:CALibration<hw>:LEVel:LOOPgain:STATe
<State>
The command switches on or off use of external loopgain correction data.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
1
Example:
CAL:LEV:LOOP:STAT ON
switches on use of external loopgain correction data.
Manual operation:
See
"Use ext. Loop Gain Correction Data"
:CALibration<hw>:LEVel[:MEASure]?
[<Force>]
Starts all adjustments which affect the level.
Query parameters:
<Force>
string
*RST:
force
Return values:
<Measure>
0 | 1
Example:
CAL:LEV:MEAS?
starts adjustments for maximum level accuracy.
Response: "0"
adjustment has been performed successfully.
Usage:
Query only
Manual operation:
See
:CALibration<hw>:LEVel:EXTern:DATA
<Data>
Queries what data has been used for the level calibration.
By default the instrument uses correction data obtained in the factory before delivery.
In addition, customer data can be used for external level correction. The customer data
is obtained using a R&S NRP power sensor. External level correction is a protected
function (see service manual, chapter 2, "Adjustment").
Parameters:
<Data>
FACTory | CUSTomer
*RST:
FACTory
Example:
CAL:LEV:EXT:DATA FACT
selects the use of the data aquired at the factory for external
level correction.
CALibration Subsystem