Instrument Function
R&S
®
SMF100A
167
Operating Manual 1167.2319.02 ─ 12
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:RETRace
Step Lin/Log - Frequency Sweep
Sets the step width for the individual frequency sweep steps.
At each step this value is added to the current frequency.
Depending on the
mode you have set, the corresponding
parameter is displayed.
"Step Lin"
The step width is a constant value in Hz.
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear]
"Step Log"
The step width is determined logarithmically in %, that means as a
constant fraction of the current frequency.
Successive frequencies are calculated as follows:
●
start_f < stop_f
f2 = f1 * (1 + step_log / 100)
If f2 > stop_f: f2 is set to stop_f.
●
start_f > stop_f
f2 = f1 / (1 + step_log / 100)
If f2 < stop_f: f2 is set to stop_f.
When the shape "Triangle" is set, the frequency values on the slope
from
stop_f back to start_f
are the same as on the slope from
start_f to stop_f
.
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic
Count - RF Frequency Sweep
Sets the number of sweep steps.
RF Frequency Block