Instrument Function
R&S
®
SMF100A
246
Operating Manual 1167.2319.02 ─ 12
To open the "User Correction" menu, select "Level Control > Configure > User Correc-
tion" or use the MENU key under "Level Control".
The combined menu "RF Level/EMF/ALC/UCOR" is divided into the several sections.
User Correction Settings
The "User Correction" settings are set in the most lower section of the combined dia-
log; this section is used to activate/deactivate user correction, and to create, select and
activate the lists.
State - User Correction
Activates/deactivates user correction.
The "UCOR" status message appears in the frequency and level display.
Remote command:
[:SOURce<hw>]:CORRection[:STATe]
User Correction Value - User Correction
Indicates the current value for level correction.
Remote command:
[:SOURce<hw>]:CORRection:VALue?
[:SOURce<hw>]:CORRection:PMETer:TYPE
Measure – User Correction
Triggers the external measurement of the user correction of the level according to the
user defined frequencies.
Remote command:
[:SOURce<hw>]:CORRection:MEASure
User Cor. Data - User Correction
Calls the "File Select" menu for selecting and creating a list or the "File Manager".
Remote command:
[:SOURce]:CORRection:CSET:CATalog?
[:SOURce<hw>]:CORRection:CSET[:SELect]
[:SOURce]:CORRection:CSET:DELete
Edit User Cor. Data - User Correction
Calls the editor for editing the selected user correction list.
RF Level