Instrument Function
R&S
®
SMF100A
124
Operating Manual 1167.2319.02 ─ 12
LAN Reset
Initiates the network configuration reset mechanism for the instrument and resets the
hostname, MAC address, and IP address.
According to the LXI standard, a LAN Reset must place the following network settings
to a default state:
Parameter
Value
TCP/IP Mode
DHCP + Auto IP Address
Dynamic DNS
Enabled
ICMP Ping
Enabled
Password for LAN configuration
LxiWebIfc
The LAN Reset for the R&S
SMF also resets the following parameters:
Parameter
Hostname
Instrument-specific host name
Description
microwave signal generator
Negotiation
Auto Detect
VXI-11 Discovery
Enabled
The LAN settings are configured using the instrument's LXI Browser Interface descri-
bed in
Chapter 2.8, "LXI Configuration"
To open the "Instrument Home Page" (welcome page), type the instrument's computer
name (host name) or IP address in the address field of the browser on your PC, for
example
http://10.111.10.175
.
Note:
Do not add the missing zeros in the IP address, while opening the Instrument
Home Page.
Remote command:
n.a.
5.2.3.18
Protection
This "Protection" dialog provides access to the unlocking of different protection levels .
Access:
► Select "Setup > Protection"
After power on the instrument, all protection levels are locked. To unlock the protec-
tion, the correct password must be entered, see
"To unlock or lock a protection level..."
General Instrument Settings