![Rohde & Schwarz SMF-B1 Operating Manual Download Page 432](http://html1.mh-extra.com/html/rohde-and-schwarz/smf-b1/smf-b1_operating-manual_1477740432.webp)
Remote Control Commands
R&S
®
SMF100A
417
Operating Manual 1167.2319.02 ─ 12
Example:
SENS:SWE:HCOP:FILE:AUTO:YEAR?
queries the year of the date part in the automatic file name.
Usage:
Query only
Options:
Option R&S SMF-K28
Manual operation:
See
" Save Options - Power Analysis "
:SENSe[:POWer]:SWEep:HCOPy:FILE[:NAME]:AUTO[:FILE]:YEAR:STATe
<State>
Activates the usage of the year in the automatic file name.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
1
Example:
SENS:SWE:HCOP:FILE:AUTO:YEAR:STAT OFF
deactivates the usage of the year in the automatic file name.
Options:
Option R&S SMF-K28
Manual operation:
See
" Save Options - Power Analysis "
:SENSe[:POWer]:SWEep:HCOPy:FILE[:NAME]:AUTO:STATe
<State>
Activates/deactivates automatic naming of the hardcopy files.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
1
Example:
SENS:SWE:HCOP:FILE:AUTO:STAT OFF
deactivates automatic naming of the hardcopy files. The file
name and directory is now defined with command
SENS:SWE:HCOP:FILE:NAME <path>
.
Options:
Option R&S SMF-K28
Manual operation:
See
" File name - Power Analysis "
:SENSe[:POWer]:SWEep:INITiate
Starts the power analysis with NRP power sensor.
Example:
SENS:SWE:INIT
start the power measurement.
Usage:
Setting only
Options:
Option R&S SMF-K28
Manual operation:
See
Power Sensor Measurement Subsystems