Remote Control Commands
R&S
®
SMF100A
540
Operating Manual 1167.2319.02 ─ 12
[:SOURce<hw>]:PULM:SOURce
<Source>
Selects the source for pulse modulation.
Parameters:
<Source>
INTernal | EXTernal | CODer | RANDom | INTernal | EXTernal |
RANDom
EXTernal
The signal applied externally via the PULSE IN connector is
used for the pulse modulation.
INTernal
The internal pulse generator is used for the pulse modulation.
Note:
This feature requires option R&S SMF-K23.
RANDom
Selects random pulses generated by the noise generator as the
source for pulse modulation.
*RST:
INTernal
Example:
PULM:SOUR EXT
selects the external modulation source.
PULM:STAT ON
activates the pulse modulation.
Options:
R&S SMF-K3 (Narrow Pulse Modulation)
Manual operation:
See
[:SOURce<hw>]:PULM:STATe
<State>
Activates the pulse modulation.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
PULM:STAT ON
activates pulse modulation.
Options:
R&S SMF-K23 (Pulse Generator)
Manual operation:
See
[:SOURce<hw>]:PULM:SYNC
<Sync>
When enabled, the external modulation signal is synchronized to the internal 200MHz
reference before it is passed to the pulse modulator. Since most instrument capabilities
are timed by the same reference, the edges of the modulation signal can be synchron-
ized e.g. to RF or to pulse modulator outputs if instrument is set up accordingly.
When disabled, the signal is passed asynchronously to the modulator.