[:SOURce<hw>]:SWEep[:FREQuency]:MARKer:XFER
Copies the frequency value of marker 1 to the start frequency of the sweep and the
frequency value of marker 2 to the stop frequency of the sweep.
The command is an event and thus has no *RST-value and no query form.
Dependencies:
Sets the start and stop frequency of the sweep.
Example:
SWE:MARK:XFER
used the frequency values of marker 1 and 2 to set the start and
stop frequency, respectively.
Usage:
Event
[:SOURce<hw>]:SWEep[:FREQuency]:MODE
<Mode>
Sets the sweep mode.
The keyword
[:FREQuency]
can be omitted (see example). The command is then
SCPI-compliant.
Parameters:
<Mode>
AUTO | MANual | STEP
AUTO
Each trigger triggers exactly one complete sweep.
MANual
The trigger system is not active. Each frequency step of the
sweep is triggered individually, either by varying the "Current
Frequency" value using the rotary knob under manual control or
by means of a
FREQ:MAN
command under remote control. With
manual control, the frequency increases or decreases (depend-
ing on the direction of the rotary encoder) by the value specified
under
FREQ:STEP:INCRement
. With remote control, the fre-
quency is set directly with the command
:FREQ:MAN
.
STEP
Each trigger triggers one sweep step only (Mode Single Step).
The frequency increases by the value entered under
SOUR:SWE:FREQ:STEP:LIN
(linear spacing) or
...:STEP:
LOG
(logarithmic spacing).
*RST:
AUTO
Example:
SWE:MODE AUTO
selects
Mode Auto
for a frequency sweep at the RF output.
Manual operation:
See
SOURce Subsystem