Instrument Function
R&S
®
SMF100A
168
Operating Manual 1167.2319.02 ─ 12
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:POINts
Sweep Time - RF Frequency Sweep
This feature is available for ramp spacing only.
Sets the sweep time for ramp sweeps. Ramp sweeps are selected above under
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:TIME
Dwell Time - Frequency Sweep
Sets the dwell time. The dwell time determines the duration of the individual sweep
steps.
The "Dwell Time" set by the user is used as the step time of the sweep. The effective
net dwell time is shorter, reduced by the setting time. This setting time may be greater
than the time specified in the data sheet.
Note:
It is recommended to switch off the display update for optimum sweep performance
especially with short dwell times (see
Chapter 5.2.3.8, "Display Update"
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:DWELl
Ext. Trigger Input Slope
Sets the polarity of the active slope of an externally applied instrument trigger.
This setting affects the TRIGGER input (BNC connector at the rear of the instrument or
the AUX connector).
"Positive"
activates the rising edge of the trigger signal.
"Negative"
activates the falling edge of the trigger signal.
Remote command:
Edit Marker
Opens a dialog used for setting and editing the markers.
For a detailed description, see
●
●
Chapter 5.6.1, "LF Gen./Noise"
RF Frequency Block