Remote Control Commands
R&S
®
SMF100A
554
Operating Manual 1167.2319.02 ─ 12
Example:
ROSC:SOUR INT
Selects the internal source.
ROSC:ADJ ON
Activates use of a user-defined adjustment value.
ROSC:ADJ:VAL 1400
Sets the adjustment value to 1400.
Manual operation:
See
[:SOURce]:ROSCillator:SOURce
<Source>
Selects the reference frequency source.
Parameters:
<Source>
INTernal | EXTernal | ELOop
INTernal
The internal reference oscillator is used.
The reference signal is provided at the REF OUT connector
EXTernal
An external reference signal is used. It must be input at the REF
IN connector at the rear of the instrument.
The instrument is informed of the frequency of the external refer-
ence signal by means of the command
ROSCillator:EXTernal:FREQuency
*RST:
---
Example:
ROSC:SOUR EXT
Selects the external source.
ROSC:EXT:FREQ 5 MHz
Informs the instrument that the external reference has a fre-
quency of 5 MHz.
Manual operation:
See
[:SOURce]:ROSCillator:OUTPut:SOURce
<Source>
The command selects the source for the reference oscillator signal output. The exter-
nal output is only available if the external reference frequency is 10 MHz.
Parameters:
<Source>
INTernal | EXTernal
INTernal
The internal reference oscillator signal is output at the REF OUT
connector.
EXTernal
The external reference oscillator signal is output at the REF
OUT connector
*RST:
---
SOURce Subsystem