Remote Control Commands
R&S
®
SMF100A
500
Operating Manual 1167.2319.02 ─ 12
Example:
SOUR:LFO2:SHAP:TRAP:PER 100s
sets a period of 100 s for the trapezoidal signal of the LF gener-
ator 2.
Options:
R&S SMF-B20
Manual operation:
See
" Period (Pulse, Triangle, Trapezoid Shape) – LF Generator
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRAPeze:RISE
<Rise>
Selects the rise time for the trapeze shape of the LF generator.
Parameters:
<Rise>
float
Range:
see data sheet
Increment: see data sheet
*RST:
10 µs
Example:
SOUR:LFO2:SHAP:TRAP:RISE 1ms
selects a rise time of 1 ms for the trapezoidal signal of the LF
generator 2.
Options:
R&S SMF-B20
Manual operation:
See
" Trapezoid Rise – LF Generator "
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRIangle:PERiod
<Period>
Sets the period of the generated pulse. The period determines the repetition frequency
of the internal signal.
Parameters:
<Period>
float
Range:
1E-6 to 100
Increment: 10E-9
*RST:
0.001
Example:
SOUR:LFO2:SHAP:TRIA:PER 100s
sets a pulse period of 100 s for the pulse shape of the signal of
LF generator 2.
Options:
R&S SMF-B20
Manual operation:
See
" Period (Pulse, Triangle, Trapezoid Shape) – LF Generator
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRIangle:RISE
<Rise>
Selects the rise time for the triangle shape of the LF generator.
SOURce Subsystem