Instrument Function
R&S
®
SMF100A
230
Operating Manual 1167.2319.02 ─ 12
Configure Pulse Data
For R&S NRP-Z power sensors that support time domain analysis and automatic pulse
analysis.
The power sensors enable pulse data analysis in measurement mode time. All impor-
tant pulse parameters are measured after setting the threshold levels. The following
graph shows most of these parameters:
The sensor calculates the pulse parameters from each measurement and delivers the
results to the R&S
SMF.
The "Pulse Data" button opens the dialog to configure and enable pulse data analysis:
●
The "Pulse Data NRP-Z- Analysis" dialog is divided into several sections. The
upper section provides the trace and the pulse data analysis. For information on
traces and the measurement data on traces, refer to
●
The "Thresholds" section covers the thresholds for detecting time values of a
pulsed signal. Thresholds are used to calculate the time parameters. The thresh-
olds can either be related to power or voltage. For information on parameters and
terms, refer to
" Voltage / Power Related - Pulse Data Analysis "
Mesial - Pulse Data Analysis "
" Proximal - Pulse Data Analysis "
●
The "Notifications" section covers the parameters "Duty Cycle", "Pulse Width",
"Pulse Period" and "Pulse Off Time", see
" Notifications - Pulse Data Analysis "
●
The "Transition Times" section covers "Rise Time", "Pulse Start Time", "Overshoot"
for the rising or falling edges, "Fall Time" and "Pulse Stop Time", see
●
The "Signal Power" section covers "Minimal Power", "Peak Power" and "Average
Power", see
" Signal Power - Pulse Data Analysis "
RF Level