Remote Control Commands
R&S
®
SMF100A
389
Operating Manual 1167.2319.02 ─ 12
Manual operation:
See
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:STOP
<Stop>
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:STARt
<Start>
Sets the start time of the selected gate. Insert value and unit.
Parameters:
<Start>
float
Increment: 1E-12
*RST:
Start/Stop: 5/15 (Gate1), 25/35 (Gate2)
Example:
CALC:SWE:TIME:GATE2:STAR 20us
sets a start time of 20 us for gate 2.
Example:
CALC:SWE:TIME:GATE2:STOP 30us
sets a stop time of 30us for gate 2.
Options:
Option R&S SMF-K28
Manual operation:
See
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:STATe
<State>
Activates the gate settings for the selected trace. The measurement ist started with
command
SENS:POW:INIT
. Both gates are active at one time.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
CALC:SWE:TIME:GATE:STAT ON
'enables time gated measurement.
Options:
Option R&S SMF-K28
Manual operation:
See
:CALCulate[:POWer]:SWEep:FREQuency:MATH<ch>:STATe
<State>
Activates the trace mathematics mode for "Frequency" measurement. This feature
enables to calculate the difference between the measurement values of two traces.
Additionally, for further calculation a math result can also be assigned to a trace.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
CALC:POW:SWE:FREQ:MATH2:STATe
Example:
switches on math mode in trace 2.
Options:
R&S SMF-K28
Manual operation:
See
" Mathematics - Trace Power Analysis "
Power Sensor Measurement Subsystems