Remote Control Commands
R&S
®
SMF100A
537
Operating Manual 1167.2319.02 ─ 12
Parameters:
<Delay>
float
Range:
10 ns to 100 s
Increment: 5 ns
*RST:
10 ns
Example:
PULM:DEL 13 us
13 us elapse after a trigger before the first pulse is generated.
Options:
R&S R&S
SMF-K23 (Pulse Generator)
Manual operation:
See
" Pulse Delay - Pulse Generator "
[:SOURce<hw>]:PULM:DOUBle:DELay
<Delay>
Sets the delay from the start of the first pulse to the start of the second pulse.
Parameters:
<Delay>
float
Range:
10 ns to 100 s
Increment: 5 ns
*RST:
3 us
Example:
PULM:DOUB:DEL 22 us
22 us elapse between the beginning of the first pulse and the
beginning of the second pulse in double-pulse mode.
Options:
R&S
SMF-K23 (Pulse Generator)
Manual operation:
See
" Double Pulse Delay - Pulse Generator "
[:SOURce<hw>]:PULM:DOUBle:STATe
<State>
Enables/disables double pulse generation. The two pulses are generated in one pulse
period.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
PULM:DOUB:STAT ON
double-pulse mode is enabled.
Options:
R&S SMF-K23 (Pulse Generator)
[:SOURce<hw>]:PULM:DOUBle:WIDTh
<Width>
Sets the width of the second pulse in case of double pulse generation.
SOURce Subsystem