Remote Control Commands
R&S
®
SMF100A
483
Operating Manual 1167.2319.02 ─ 12
[:SOURce]:LFOutput<ch>:SOURce
<Source>
The command selects the internal and/or external source to be used for the LF Output
signal.
Parameters:
<Source>
LF1 | LF2 | NOISe | EXT1 | EXT2
*RST:
LF1
Example:
SOUR:LFO1:SOUR NOIS
selects the noise generator as source for the LF output signal.
Options:
R&S SMF-B20
Manual operation:
See
[:SOURce]:LFOutput<ch>[:STATe]
<State>
Activates the LF signal output.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
OFF
Example:
SOUR:LFO1 ON
activates the LF output.
Options:
R&S SMF-B20
Manual operation:
See
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:DWELl
<Dwell>
Sets the dwell time for each frequency step of the sweep.
Tip:
It is recommended to switch off the "Display Update" for optimum sweep perform-
ance especially with short dwell times (
Parameters:
<Dwell>
float
Range:
see data sheet: Dwell time setting range
Increment: 100E-6
*RST:
15E-3
Example:
LFO:SWE:DWEL 20 ms
sets a dwell time of 20 ms.
Options:
R&S SMF-B20
Manual operation:
See
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:EXECute
Immediately starts an LF frequency sweep in
.
SOURce Subsystem