Remote Control Commands
R&S
®
SMF100A
486
Operating Manual 1167.2319.02 ─ 12
Options:
R&S SMF-B20
Manual operation:
See
" State - LF Frequency Marker "
[:SOURce]:LFOutput:SWEep[:FREQuency]:MARKer<ch>:VSTate
<VState>
The command activates or deactivates the selected amplitude marker. If activated, the
level is reduced by the amplitude entered with command
SWEep[:FREQuency]:MARKer:AMPLitude
on reaching the mark.
Parameters:
<VState>
0 | 1 | OFF | ON
*RST:
0
Example:
SOUR:LFO:SWE:FREQ:MARK1:VST ON
the output level is reduced by a constant value if frequency
marker 1 is executed.
Options:
R&S SMF-B20
Manual operation:
See
" Amplitude Marker - LF Frequency Marker "
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:MODE
<Mode>
Sets the cycle mode of the LF sweep.
The assignment of the GPIB commands to the sweep modes is given in the description
of the sweep dialogs.
SOURce Subsystem