Instrument Function
R&S
®
SMF100A
285
Operating Manual 1167.2319.02 ─ 12
Up to ten frequency markers can be defined. The frequency of the marker signal at the
MARKER output changes according to the selected polarity when the frequency mark
is reached. For each of these markers an amplitude marker can be activated in addi-
tion. If activated, the marker signal is attenuated by the set amplitude when the fre-
quency mark is reached. Only one marker is output at the MARKER output at one time.
Marker - LF Frequency Marker
Displays the marker index.
Remote command:
n.a.
Frequency - LF Frequency Marker
Enters the frequency for the corresponding marker. If the marker is activated, the sig-
nal level at the MARKER output changes on reaching this frequency mark.
Remote command:
[:SOURce]:LFOutput:SWEep[:FREQuency]:MARKer<ch>:FREQuency
State - LF Frequency Marker
Switches in or off the selected marker. If on, the marker is output at the MARKER out-
put on reaching the mark.
Remote command:
[:SOURce]:LFOutput:SWEep[:FREQuency]:MARKer<ch>:FSTate
Amplitude Marker - LF Frequency Marker
Activates or deactivates the selected amplitude marker. If activated, the level is
reduced by the amplitude entered below on reaching the mark.
Remote command:
[:SOURce]:LFOutput:SWEep[:FREQuency]:MARKer<ch>:VSTate
Amplitude - LF Frequency Marker
Sets the amplitude marker attenuation.This value is valid for all markers.
Remote command:
[:SOURce]:LFOutput:SWEep[:FREQuency]:MARKer:AMPLitude