Instrument Function
R&S
®
SMF100A
226
Operating Manual 1167.2319.02 ─ 12
Indicates the type of R&S NRP power sensor assigned to the selected trace. This field
is automatically updated if the sensor is connected or disconnected. Additionally, this
sensor is indicated on the trace button in the measurement diagram.
Remote command:
n.a.
State - Gate
Enables time gated measurement. The measurement is started with the "Start" button
in the main measurement diagram. Both gates are active at one time.
The gate borders and the measurement values (average and peak power) are indica-
ted in/below the measurement diagram. The indication is only available for certain dia-
gram views which are switched with the "Rearrange" key.
Remote command:
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:STATe
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:AVERage?
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:MAXimum?
:TRACe[:POWer]:SWEep:MEASurement:GATE:DISPlay:ANNotation[:STATe]
Start / Stop - Gate
Sets the start and the stop times for the respective gate.
Remote command:
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:STARt
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:STOP
Diagram... - Power Analysis
Returns to the "NRP-Z Analysis" diagram.
Configure Sensors
The "Configure NRP-Z Analysis" dialog is divided into several sections. The power
sensor area additionally provides specific parameters for the power sensor. This part of
the dialog can differ from the following description depending on the sensor used.
Refer to the manual of the power sensor in this case.
The "Pulse Data..." button is displayed only for measurement mode time and if an R&S
NRP-Z81 power sensor is connected.
RF Level